A microdestructive capillary electrophoresis method for the analysis of blue-pen-ink strokes on office paper
Identifiers
Permanent link (URI): http://hdl.handle.net/10017/50108DOI: 10.1016/j.chroma.2015.04.036
ISSN: 0021-967
Date
2015-06-26Affiliation
Universidad de Alcalá. Departamento de Química Analítica, Química Física e Ingeniería QuímicaBibliographic citation
Journal of Chromatography A., 2015, v. 1400, p. 140-148
Keywords
Blue ink
blue pen
capillary electrophoresis
dyes
ink differentiation
profiling
Document type
info:eu-repo/semantics/article
Version
info:eu-repo/semantics/publishedVersion
Rights
Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)
© Elsevier, 2015
Access rights
info:eu-repo/semantics/openAccess
Abstract
This manuscript describes the development of a capillary electrophoresis (CE) method for the detection of acid and basic dyes and its application to real samples, blue-pen-ink strokes onoffice paper. First, a capillary zone electrophoresis (CZE) method was developed for the separation of basic and acid dyes, by studying the separation medium (buffer nature, pH andrelative amount of additive) and instrumental parameters (temperature, voltage and capillary dimensions). The method performance was evaluated in terms of selectivity, resolution (above 5 and 2 for acid dyes and basic dyes, respectively, except for two basic dye standards), LOD (lower than 0.4 mg/L) and precision as intraday and interday RSD values of peak migrationtimes (lower than 0.6 %). The developed method was then applied to 34 blue pens from different technologies (rollerball, ballpoint, markers) and with different ink composition (gel, 30 water-based, oil-based). A microdestructive sample treatment using a scalpel to scratch 0.3 mg of ink stroke was performed. The entire electropherogram profile allowed the visual discrimination between different types of ink and brands, being not necessary a statistical treatment. A 100% of discrimination was achieved between pen technologies, brands, and models, although non-reproducible zones in the electropherograms were found for blue gel pen samples. The two different batches of blue oil-based pens were also differentiated. Thus, this method provides a simple, microdestructive, and rapid analysis of different blue pentechnologies which may complement the current analysis of questioned documents performed by forensic laboratories.
Files in this item
Files | Size | Format |
|
---|---|---|---|
Microdestructive_Calcerrada_JC ... | 440.7Kb |
![]() |
Files | Size | Format |
|
---|---|---|---|
Microdestructive_Calcerrada_JC ... | 440.7Kb |
![]() |
Collections
- QUANING - Artículos [257]